CPX Specifications

Sample cooling assembly 

Type:Type: 1 + 3-stage split flow cryostat (sample + 3 radiation shields); dual exhaust ports with independent flow control of sample and shields
Radiation shielding:Inner 4 K (partial) radiation shield; main radiation shield with top viewport surrounds sample; outer second (partial) radiation shield
Cooling source:Liquid helium (LHe) or liquid nitrogen (LN2)
  

Magnetic field:

Ring magnet kit

Temperature1

Base temperature Control range Configuration
4.3 K4.4 K to 475 KUp to 6 probe arms installed
1.9 K2 K to 475 Kwith PS-LT option
1.6 K 1.65 K to 400 Kwith PS-VLT-CPX option
10 K10 K to 400 Kwith PS-LL-CPX option
20 K20 K to 675 Kwith PS-HTSTAGE option

1Base temperature assumes heat load for probes thermally anchored as listed in probing section

Control stability

Sample stage temperature Helium stability Nitrogen stability
Base temperature (no heater control)±15 mK±20 mK
<10 K±50 mK
10 K to 100 K±20 mK±50 mK
101 K to 250 K±15 mK±50 mK
251 K to 350 K±15 mK±100 mK
351 K to 475 K±50 mK±100 mK
476 K to 675 K±50 mK±100 mK

Vacuum

Standard TPS-FRG PS-HV-CPX option
Pump down time30 min
(to reach <1 × 10-3 Torr)
10 min
(to reach <1 × 10-3 Torr)
Room temperature<5 × 10-4 Torr<5 × 10-6 Torr
Helium base temperature<1 × 10-5 Torr<5 × 10-7 Torr
Maximum temperature<5 × 10-3 Torr<5 × 10-7 Torr

Cycle time (LHe):

3 h
Vacuum pump down:0.5 h
Station cooldown:1.25 h
Station warmup:1.25 h
Load-lock sample exchange:1 h with PS-LL-CPX option

Vibration at sample

Standard PS-PVIS option
X axis<300 nm (11.8 µin)<30 nm (1.18 µin)
Y axis
Z axis

Sample

 
Maximum sample size:51 mm (2 in) diameter
25.4 mm (1 in) diameter with PS-VLT-CPX option
25.4 mm (1 in) diameter with PS-LL-CPX option
Sample backside optical access:Not an option
Sample rotation:±5° in-plane rotation

Probing configurations

 
Maximum probes:6
Probe arm sensor:Included for monitoring probe temperature
Cooled probe mounts:<20 K with sample at base temperature
Probe mount:Thermally anchored to sample stage or 4 K shield
Probe arm:Thermally anchored to radiation shield
DC/RF probes and DC/RF flexible probes:Electrical isolation >100 GΩ for low leakage measurements
Microwave probes:Frequency range DC up to 67 GHz for RF measurements
Fiber optic probes:Available for electro-optical measurements
Probe landing:All probes can land at single point in a 25.4 mm (1.00 in) diameter circle

Cryogen consumption2

Room to base temperature (total):10 to 15 liquid L helium; 4 to 5 liquid L nitrogen
Helium at 5 K:1 liquid L/h
Nitrogen at 80 K:<0.5 liquid L/h

2Assumes maximum number of probe arms installed

Station details

Temperature equipment

Sample stage sensor/heater power:DT-670-CU-HT calibrated silicon diode/50 W
(PS-VLT sensor optimized for <4 K)CX-1050-CU-HT-1.4M calibrated Cernox/50 W
4 K radiation shield sensor/heater power:DT-670C-CU silicon diode/100 W
Radiation shield sensor/heater power:DT-670C-CU silicon diode/100 W
2nd radiation shield sensor/heater power:DT-670C-CU silicon diode/50 W
Probe arm:DT-670C-SD silicon diode (on one probe arm)/no active control
Electronics console:
Temperature controllers:Two Model 336s
Helium transfer line:Large bore helium/nitrogen transfer line with flow control 12.7 mm (0.5 in) top port withdrawal
  

Included sample holder:

Removable SH-1.25-G, 31.8 mm (1.25 in) maximum sample size, grounded
  

Station footprint:

800 mm (31.5 in) width × 670 mm (26.4 in) depth
(does not include instrument console)
  

Vacuum chamber

Material:Electroless nickel-plated aluminum
Overall size:280 mm (11 in) outer diameter × 343 mm (13.5 in) tall
Chamber access:229 mm (9 in) diameter opening
Viewport:63.5 mm (2.50 in) diameter fused quartz
Viewing area:54.0 mm (2.13 in) diameter
Pump port:NW 40 with included vacuum isolation valve
Gauge port:NW 25 (two ports)
Gas port:0.25 in NPT with valve
Over pressure safety:3.5 kPa (0.5 psi) pressure relief valve
  

Radiation shielding

Material:Electroless nickel-plated aluminum
Sample access:167 mm (6.6 in) diameter lid
Viewport:50 mm (1.97 in) diameter IR absorbing window
Viewing area:50 mm (1.97 in) diameter
  

Vision system

  

Included kits

Tool kit:Included tools for standard operation
Spares kit:Included kit for common consumable items
Probe starter kit:ZN50R-25-W (qty 2), ZN50R-25-BECU (qty 2), and ZN50R-CVT-25-W (qty 2) for training
  

Utilities

1-phase voltage:100/120/220/240 VAC (+5%, -10%), 50/60 Hz
1-phase power: 2 kVA recommended
3-phase voltage:Not required
3-phase power:Not required
Cooling water:Not required
  

Approval:

All instruments CE marked


 

Back to the CPX overview

Probing configurations

Vision system

This website uses cookies to improve user experience. By using our website you consent to all cookies in accordance with our Privacy Policy.   More info