Tangential 400 Series Hall Probe Specifications

Click here for Hall probe active areas

By default, each probe has a 2 m long cable (except for Model 410 and cryogenic probes). Also available: 6 m, 10 m, and 30 m lengths.

Tangential probes

Tangential 400 Series Hall probe

Operating temperature range: 0 °C to +75 °C

 RoHS compliant
L mm (in)T mm (in)W mm (in)A mm (in)Active area mm (in)Stem materialFrequency rangeUsable full scale rangesCorrected accuracy (% rdg at 25 °C) Temp coefficient (max) zeroTemp coefficient (max) of sensitivityContains temp sensor
for Models 475, 455, and 425
HMNTAN-DQ02-THNo38.1 ±1.52 (1.5 ±0.060)2.8 (0.11) max6.6 (0.26) max0.76 ±0.18 (0.030 ±0.005)Approx 0.51 dia (0.020 dia) PlasticDC to 400 HzHSE
3.5 G, 35 G, 350 G, 3.5 kG, 35 kG
0.25% to 20 kG±0.1 G/°C455/475: ±0.02%/°C

425: -0.05%/°C
Yes
 RoHS compliant
L mm (in)T mm (in)W mm (in)A mm (in)Active area mm (in)Stem materialFrequency rangeUsable full scale rangesCorrected accuracy (% rdg at 25 °C)Temp coefficient (max) zeroTemp coefficient (max) of sensitivityContains temp sensor
for Models 460, 450, and 421
MNTAN-DQ02-THNo38.1 ±1.52 (1.5 ±0.060)2.8 (0.11) max6.6 (0.26) max0.76 ±0.18 (0.030 ±0.005)Approx 0.51 dia (0.020 dia) PlasticDC and 10 Hz to 400 HzHSE-1
30 G, 300 G, 3 kG, 30 kG
0.25% to 20 kG±0.1 G/°C-0.05%/°CNo