11 Control stability of the electronics only, in ideal thermal system12 Control source error has negligible effect on measurement accuracy13 Diode input excitation current can be set to 1 mA — refer to the Model 325 user manual for details14 Current source error is removed during calibration 15 Accuracy specification does not include errors from room temperature compensation
Number of inputs
2
Input configuration
Each input is factory configured as either diode/RTD or thermocouple
Isolation
Sensor inputs optically isolated from other circuits but not from each other
A/D resolution
24-bit
Input accuracy
Sensor dependent – refer to Input Specifications table
Measurementresolution
Maximum update rate
10 rdg/s on each input (except 5 rdg/s on input A when configured as thermocouple)
User curves
Room for 15 200-point CalCurves™ or user curves
SoftCal™
Improves accuracy of DT-470 diode to ±0.25 K from 30 K to 375 K; improves accuracy of Platinum RTDs to ±0.25 K from 70 K to 325 K; stored as user curves
Filter
Averages 2 to 64 input readings
Control loops
Control type
Closed-loop digital PID with manual heater output or open loop
Tuning
Autotune (one loop at a time), PID, PID zones
Control stability
Sensor dependent – see Input Specification table
PID control parameters
Proportional (gain)
0 to 1000 with 0.1 setting resolution
Integral (reset)
1 to 1000 (1000/s) with 0.1 setting resolution
Derivative (rate)
1 to 200% with 1% resolution
Manual output
0 to 100% with 0.01% setting resolution
Zone control
10 temperature zones with P, I, D, manual heater out, and heater range
Setpoint ramping
0.1 K/min to 100 K/min
Safety limits
Curve temperature, power up heater off, short-circuit protection
No. of reading displays
1 to 4
Display units
K, ºC, V, mV and Ω
Reading source
Temperature, sensor units
Display update rate
2 rdg/s
Temp display resolution
0.001º from 0º to 99.999º, 0.01º from 100º to 999.99º, 0.1º above 1000º
Sensor units displayresolution
Sensor dependent; to 5 digits
Other displays
Setpoint, Heater Range, and Heater Output (user selected)
Setpoint setting resolution
Same as display resolution(actual resolution in sensor dependent)
Heater output resolution
1%
Display annunciators
Control input, Remote, Autotune
Keypad
20-key membrane, numeric and specific functions
Front panel features
Front panel curve entry, keypad lock-out
Features
SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1
Reading rate
To 10 rdg/s on each point
Software support
LabVIEW™ driver
Electrical format
Baud rates
9600, 19200, 38400, 57600
Connector
9-pin D-style, DTE configuration
To 10 rdgs/s on each input
Power requirement
100, 120, 220, 240 VAC, +6%, -10%, 50 or 60 Hz, 85 VA
Size
216 mm W × 89 mm H × 368 mm D (8.5 in × 3.5 in × 14.5 in), half rack
Weight
4.00 kg (8.82 lb)
Approval
CE mark