STVP-100-FTIR Cryostat for FTIR Spectroscopy
Optical continuous flow cryostats for FTIR applications. These cryostats are similar to our standard model STVP-100 cryostats, but with the modifications for integrating the cryostat with your spectrometer system.
Standard configuration

Similar in operation to the standard SuperTran-VP (STVP-100) cryostats, the STVP-100-FTIR includes a multiple position sample holder and a mounting flange to allow integration with a spectrometer. The sample is cooled by flowing vapor and is therefore an ideal choice for studying materials that are difficult to cool by conduction.
The mechanical drawing shows a standard STVP-100 sample positioner with coarse translation and rotation. If more precise translation and/or rotation is required, a linear manipulator may be mounted on the sample positioner for more accurate positioning. The drawing illustrates typical Fourier Transform Infra-Red (FTIR) windows, which may be altered to suit the specific application. The sample tube diameter may also be specified.
Standard STVP-100-FTIR features include:
- Sample in flowing vapor
- Multiple position sample holder
- Adjustable position sample positioner allowing sample rotation and translation
- Silicon diode temperature sensors on sample mount and vaporizer
- Mounting flange to fit spectrometer
- High-efficiency cryogen transfer line
- Indium sealed, strain-relieved, cold windows
- Liquid helium cooling
Options include:
- Manually controlled precision linear and/or rotary manipulators
- Remotely controlled linear manipulator and/or rotary stage
- IR window materials, wedged or plane parallel
- Range of sample tube sizes available
- Liquid nitrogen cooling
- Electrical access to the sample
- A complete system solution including temperature controllers, pumping systems, and storage Dewars
- Isothermal zone sample tube (provides uniform sample temperature and enhanced operation with LN2)
- Base plate or mounting flange to interface with a customer-owned spectrometer
Contact us today!

STVP-100-FTIR Specifications
Sample environment | Liquid or flowing vapor |
Temperature range | <2 to 300 K |
Initial cooldown time | 15 min |
Temperature stability (with controller) | 50 mK or less |
Orientation | Vertical for <4.5 K |
System weight (less transfer line) | Approximately 15 lb |
Cryogen consumption (cooldown) | 0.5 L LHe from 325 K to 4.2 K |
Cryogen consumption (operation) | 1.1 L/h at 5 K with LHe |
NOTE: Specifications do not include parasitic or experimental heat loads. |
Optional configurations
STVP-100-FTIR with mounting flange designed for use in a Bomem FTIR spectrometer
This cryostat was designed for use in a Bomem FTIR spectrometer. It has ZnSe outer windows in one axis and polyethylene windows in the other axis. The user requested a precision sample positioner with a special compact three-position sample holder and a manual precision linear/rotary manipulator.